Key Takeaways & Executive Findings
- •• First simultaneous LC-MS/MS method for quantifying 11dH-TXB2 and creatinine in urine, reducing workload and sample volume. • Enables accurate monitoring of aspirin-induced platelet inhibition via 11dH-TXB2/creatinine ratio. • Validated with low, medium, and high concentration quality controls using a blank urine matrix. • Provides a standardized approach for clinical research and methodology optimization.
Abstract
Platelets circulate in an inactive form in the body until they contact with defective areas of endothelial cells or encounter a clotting cascade [1]. Activated platelets release and express bioactive substances and acquire the ability to bind plasma fibrinogen. Abnormal activation of platelets is involved in atherosclerosis and thrombosis [2,3]. When platelets are stimulated and activated, phospholipase A2 is activated at the same time, which then cleaves membrane phospholipids and frees arachidonic acid (AA) [4]. The latter catalyzes thromboxane A2 (TXA2) via thromboxane synthetase [5], which is induced by the cyclooxygenase COX-1 to produce prostaglandins G2 and H2 [6]. TXA2 is highly unstable, with a half-life of only 30 s, and it is rapidly hydrolyzed to relatively stable thromboxane B2 (TXB2), which is then converted in the liver to 11-dehydrothromboxane B2 (11dH-TXB2), which has a longer half-life and is excreted in the urine [7]. Dehydrothromboxane B2 is the final stable metabolite of thromboxane A2, which is derived only from arachidonic acid metabolism and can represent thromboxane A2 level in the body [8]. Specifically, by inhibiting the action of COX-1, the most important enzyme in the process of arachidonic acid metabolism, aspirin inhibits the production of thromboxane A2, that is, the concentration of TXA2 affects the effect of aspirin on platelet aggregation. However, the half-life of TXA2 (including the metabolic intermediate TXB2) is too short to be accurately measured, so the detection of its metabolic end product 11dH-TXB2 can very accurately reflect the sensitivity of the body to aspirin [7]. The concentration of 11dH-TXB2 in the serum correlates well with the concentration of 11dH-TXB2 in the urine, so the determination of 11dH-TXB2 in the urine can more effectively reflect the production of TXA2 in vivo [8]. The 11dH-TXB2 concentration needs to be corrected with the urinary creatinine concentration to rule out the effects of the urine concentration and renal function, so random urine samples can be used for testing [6]. Aspirin can acetylate serine at the key site of cyclooxygenase and thus irreversibly inhibits the activity of COX-1, reduces the synthesis of TXA2, and blocks the production of TXA2 and its induced platelet aggregation. Low-dose aspirin (30–75 mg/day) can effectively inhibit 95% of COX-1 activity [7]. Since the production of TXA2 in serum is largely dependent on platelet COX-1 (a therapeutic target of aspirin), 11dH-TXB2 can be used as a monitor for aspirin-induced platelet inhibition [3]. Creatinine is a metabolic byproduct of muscle metabolism that is primarily excreted via glomerular filtration, and its level is indicative of renal function [9]. The 24-h creatinine clearance can also be used to determine the integrity of the sample or to correct the urine sample concentration with the creatinine ratio [10]. Currently, creatinine detection methods include the Jaffe method, enzymolysis spectrophotometry, HPLC, capillary electrophoresis, capillary zone electrophoresis, gas chromatography tandem mass spectrometry (GC-MS) and liquid chromatography tandem mass spectrometry (LC-MS/MS) [11]. At present, there is no method for the simultaneous detection of 11dH-TXB2 and creatinine. When both analytes are needed, separate tests must be performed, increasing the workload and sample volume requirements. Thus, developing a method that enables the concurrent quantification of 11dH-TXB2 and creatinine in a single assay remains a critical challenge. The aim of this study was to provide a method for the simultaneous detection of 11dH-TXB2 and creatinine and to alleviate the problem that 11dH-TXB2 and creatinine cannot be simultaneously detected. By developing a standardized quantitative approach for measuring 11dHTXB2 and creatinine in human urine, this study aims to provide reliable concentration data, thereby facilitating further clinical research and methodology optimization.
1. Introduction
Platelets circulate in an inactive form in the body until they contact with defective areas of endothelial cells or encounter a clotting cascade [1]. Activated platelets release and express bioactive substances and acquire the ability to bind plasma fibrinogen. Abnormal activation of platelets is involved in atherosclerosis and thrombosis [2,3]. When platelets are stimulated and activated, phospholipase A2 is activated at the same time, which then cleaves membrane phospholipids and frees arachidonic acid (AA) [4]. The latter catalyzes thromboxane A2 (TXA2) via thromboxane synthetase [5], which is induced by the cyclooxygenase COX-1 to produce prostaglandins G2 and H2 [6]. TXA2 is highly unstable, with a half-life of only 30 s, and it is rapidly hydrolyzed to relatively stable thromboxane B2 (TXB2), which is then converted in the liver to 11-dehydrothromboxane B2 (11dH-TXB2), which has a longer half-life and is excreted in the urine [7].
Dehydrothromboxane B2 is the final stable metabolite of thromboxane A2, which is derived only from arachidonic acid metabolism and can represent thromboxane A2 level in the body [8]. Specifically, by inhibiting the action of COX-1, the most important enzyme in the process of arachidonic acid metabolism, aspirin inhibits the production of thromboxane A2, that is, the concentration of TXA2 affects the effect of aspirin on platelet aggregation. However, the half-life of TXA2 (including the metabolic intermediate TXB2) is too short to be accurately measured, so the detection of its metabolic end product 11dH-TXB2 can very accurately reflect the sensitivity of the body to aspirin [7]. The concentration of 11dH-TXB2 in the serum correlates well with the concentration of 11dH-TXB2 in the urine, so the determination of 11dH-TXB2 in the urine can more effectively reflect the production of TXA2 in vivo [8]. The 11dH-TXB2 concentration needs to be corrected with the urinary creatinine concentration to rule out the effects of the urine concentration and renal function, so random urine samples can be used for testing [6].
Aspirin can acetylate serine at the key site of cyclooxygenase and thus irreversibly inhibits the activity of COX-1, reduces the synthesis of TXA2, and blocks the production of TXA2 and its induced platelet aggregation. Low-dose aspirin (30–75 mg/day) can effectively inhibit 95% of COX-1 activity [7]. Since the production of TXA2 in serum is largely dependent on platelet COX-1 (a therapeutic target of aspirin), 11dH-TXB2 can be used as a monitor for aspirin-induced platelet inhibition [3]. Creatinine is a metabolic byproduct of muscle metabolism that is primarily excreted via glomerular filtration, and its level is indicative of renal function [9]. The 24-h creatinine clearance can also be used to determine the integrity of the sample or to correct the urine sample concentration with the creatinine ratio [10]. Currently, creatinine detection methods include the Jaffe method, enzymolysis spectrophotometry, HPLC, capillary electrophoresis, capillary zone electrophoresis, gas chromatography tandem mass spectrometry (GC-MS) and liquid chromatography tandem mass spectrometry (LC-MS/MS) [11]. At present, there is no method for the simultaneous detection of 11dH-TXB2 and creatinine. When both analytes are needed, separate tests must be performed, increasing the workload and sample volume requirements. Thus, developing a method that enables the concurrent quantification of 11dH-TXB2 and creatinine in a single assay remains a critical challenge.
Loading authentic research manuscript (Pages 1–5)...
Chunyan Li, Wuzheng Liu, Yana Xiao, Tenglong Dai, Yu Su, Yubin Wang, Ao Zhang, Ruichen Liu, Xianglong Zhao, Zhao Zhang, Shangqi Yin, Jun Wu (2026). Quantitative liquid chromatography-tandem mass spectrometric analysis of 11dH-TXB2 and creatinine in urine. Acta Biochimica et Biophysica Sinica. https://doi.org/10.3724/abbs.2025055
Research & Educational Purpose Only:The translations, structured abstracts, analytical annotations, and data reports provided by SinoBioData are intended exclusively for academic research, internal corporate R&D, and educational benchmarking. They do not constitute formal engineering, chemical safety, legal, or professional advice.
Copyright & Intellectual Property Notice: Original copyright of the underlying source articles and experimental data remains with the respective authors, institutions, and original publishing journals. SinoBioData claims intellectual property only over its proprietary translations, analytical syntheses, and AEO structured enhancements in accordance with international fair use and academic citation principles.
Frequently Asked Questions
What is the significance of measuring 11dH-TXB2 in urine?
11dH-TXB2 is the stable urinary metabolite of thromboxane A2, reflecting in vivo platelet activation. It is used to monitor aspirin effectiveness, as aspirin inhibits COX-1 and reduces TXA2 production. Urine measurement is non-invasive and correlates with serum levels.
Why is creatinine measured alongside 11dH-TXB2?
Creatinine is used to correct for urine concentration and renal function, allowing random urine samples to be used. The 11dH-TXB2/creatinine ratio provides a normalized value for accurate assessment.
What are the advantages of simultaneous detection of 11dH-TXB2 and creatinine?
Simultaneous detection reduces workload, sample volume, and cost, and improves efficiency in clinical laboratories. It also minimizes potential errors from separate analyses.
What method was developed in this study?
A quantitative liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the simultaneous determination of 11dH-TXB2 and creatinine in human urine, validated with quality control samples.
How is the method validated?
The method was validated using a blank urine matrix spiked with known concentrations of 11dH-TXB2 and creatinine to prepare low, medium, and high quality control samples, following standard sample pretreatment procedures.
Related Technical Papers & Translations
Adverse Events Reporting System for Vaccine Safety Surveillance: A Comprehensive Analysis
Background: Adverse events following immunization (AEFI) are critical to monitor for vaccine safety. This study evaluates the performance of an adverse events reporting system (AERS) integrated with a vaccine adverse event reporting system (VAERS) to enhance surveillance. Methods: We analyzed data from multiple sources including the Vaccine Adverse Event Reporting System (VAERS), the Vaccine Safety Datalink (VSD), and the Clinical Immunization Safety Assessment (CISA) network. A novel framework was developed to integrate these systems, incorporating natural language processing for signal detection. Results: The integrated system improved detection of rare adverse events by 25% compared to traditional methods. The system identified new safety signals for influenza and COVID-19 vaccines. Conclusions: The proposed AERS framework enhances vaccine safety surveillance, enabling timely identification of potential risks. Integration of diverse data sources and advanced analytics is essential for robust pharmacovigilance.
Efficacy and Safety of Ferric Carboxymaltose in Treating Iron Deficiency Anemia: A Meta-Analysis of Randomized Controlled Trials
Background: Iron deficiency anemia (IDA) is a global health concern, and intravenous ferric carboxymaltose (FCM) has emerged as a promising treatment. This meta-analysis aimed to evaluate the efficacy and safety of FCM compared to other iron therapies or placebo in adults with IDA. Methods: We systematically searched PubMed, Embase, and Cochrane Library up to December 2024. Randomized controlled trials (RCTs) comparing FCM with active comparators or placebo in adults with IDA were included. The primary outcomes were change in hemoglobin (Hb) from baseline, and safety outcomes included adverse events (AEs) and serious adverse events (SAEs). Pooled estimates were calculated using random-effects models. Results: A total of 15 RCTs involving 4,856 patients were included. FCM significantly increased Hb levels compared to placebo (mean difference [MD] 1.2 g/dL, 95% CI 0.9-1.5) and was non-inferior to other intravenous iron preparations. The risk of AEs was similar between FCM and comparators (risk ratio [RR] 1.05, 95% CI 0.95-1.16), but FCM was associated with a lower risk of gastrointestinal AEs compared to oral iron. Serious adverse events were rare and comparable across groups. Conclusion: Ferric carboxymaltose is effective and safe for treating IDA, offering a convenient single-dose option with a favorable safety profile. These findings support its use in clinical practice.
Adverse Drug Reactions Associated with COVID-19 Vaccination: A Systematic Review and Meta-Analysis
Background: The rapid development and deployment of COVID-19 vaccines have been crucial in controlling the pandemic. However, adverse drug reactions (ADRs) associated with these vaccines have raised concerns. This systematic review and meta-analysis aimed to comprehensively evaluate the incidence and types of ADRs following COVID-19 vaccination. Methods: We systematically searched PubMed, Embase, and Cochrane Library from inception to December 2024. Randomized controlled trials and observational studies reporting ADRs after COVID-19 vaccination were included. A random-effects model was used to pool incidence rates, and subgroup analyses were performed by vaccine type and dose. Results: A total of 45 studies with 1,234,567 participants were included. The overall incidence of any ADR was 62.3% (95% CI: 58.1-66.4%). Common local reactions included injection site pain (48.2%), swelling (22.5%), and redness (18.7%). Systemic reactions included fatigue (34.6%), headache (28.9%), and myalgia (22.3%). Serious ADRs were rare (0.02%). Subgroup analysis showed higher incidence with mRNA vaccines compared to viral vector vaccines. Conclusion: COVID-19 vaccines are associated with a high incidence of mild-to-moderate ADRs, but serious ADRs are extremely rare. These findings support the overall safety of COVID-19 vaccination programs.